HPLC EX1800LPC
With full functional configuration to meet various application needs of industries
The infusion unit is driven by a linear motor with leading technology, which significantly improves the accuracy and precision
Easily deal with complex samples by the strong separation performance of two-dimensional separations system
Professional workstation software, are complies with GMP, FDA 21 CFR, 3Q and other regulatory certification requirements
EX1800 Digital Quaternary Low Pressure Infusion Pump
Flow rate range: 0.001-10.000 mL/min, in 0.001 mL increments
Flow accuracy: ≤±0.1% (1.00ml/min, H2O,R.T.,8.5MPa);
Flow precision: RSD≤0.06% (1.00ml/min, H2O,R.T.,30MPa)
Maximum pressure: 10150 psi(70MPa);
Pressure accuracy: ≤±1.0%;
Pressure pulsation: ≤0.1 MPa
P1800 Built-in Low-Pressure Gradient Unit
Vacuum degasser, 4-6 channel
Composition rang: 0-100%;
Composition precision: ≤0.2% RSD
Composition accuracy: ≤±2.0% ;
Mixed Accuracy :±2% ;
Mixing precision :±0.3% EX18001UV/ D (dual wavelength UV detector) It can simultaneously monitor two wavelength signals, achieve the best detection of different components, and also calculate the wavelength ratio to check peak purity . Wavelength range: 190-800 nm
Wavelength repeatability: ≤0.1 nm ;
Wavelength accuracy: ≤±1.0 nm
Bandwidth: 8nm ;Response time: 0.1 s-2 s
Baseline noise: ≤±1×10-5 AU
Drift: ≤3×10-4 AU/h (empty cell, 254 nm, 60 min)
Path length: 8 mm(analytical cell)
Light source: Deuterium lamp+tungsten lamp EX-Clarity Chromatographic Data-processing Workstation
Internal and external standard calculation methods
Automatically displays, prints, exports and starts other programs after the completion of a measurement.
Comply with the requirements of the 21 CFR Part 11 directive of the FDA EX-Clarity System Kits Containing required tubing, fittings, wrenches and cables for system.
Original Series C18 5μm ID4.6×250mm
Degasser, 4 -6channel