HPLC Gradient System EX1800
With full functional configuration to meet various application needs of industries
The infusion unit is driven by a linear motor with leading technology, which significantly improves the accuracy and precision
Easily deal with complex samples by the strong separation performance of two-dimensional separations system
Professional workstation software, are complies with GMP, FDA 21 CFR, 3Q and other regulatory certification requirements
EX1800 HPLC Gradient System Flow rate range: 0.001-10.000 mL/min, in 0.001 mL increments
Flow accuracy: ≤±0.1% (1.00ml/min,); Flow precision: RSD≤0.06% (1.00ml/min,)
Maximum pressure: 9000 psi(62MPa) ; Pressure accuracy: ≤±1.0%; Pressure pulsation: ≤0.1 MPa P1800 Binary High Pressure Gradient System
Flow Presion: 0RSD<0.06%; Fluctuation: 0.1MPa ; Power Supply: 110V / 220V ; EX1800 SM High pressure gradient mixer Mixed volume :80µL
Built-in Low-Pressure Gradient Unit degasser, 2 channel
Composition rang: 0-100%; Composition precision: ≤0.2% RSD
Composition accuracy: ≤±2.0% ; Mixed Accuracy :±2% ; Mixing precision :±0.3% EX18001UV/ D (dual wavelength UV detector) It can simultaneously monitor two wavelength signals, achieve the best detection of different components, and also calculate the wavelength ratio to check peak purity . Wavelength repeatability: ≤0.1 nm ; Wavelength accuracy: ≤±1.0 nm
Bandwidth: 8nm ;Response time: 0.1 s-2 s
Baseline noise: ≤±1×10-5 AU
Drift: ≤3×10-4 AU/h (empty cell, 254 nm, 60 min)
Path length: 8 mm(analytical cell)
Light source: Deuterium lamp+tungsten lamp
EX-Clarity Chromatographic Data-processing Workstation
Internal and external standard calculation methods
Automatically displays, prints, exports and starts other programs after the completion of a measurement.
Comply with the requirements of the 21 CFR Part 11 directive of the FDA EX-Clarity System Kits Containing required tubing, fittings, wrenches and cables for system.
Original Series C18 5μm ID4.6×250mm