HPLC Gradient System LC100-HPLC V3
Reliable and stable, visible touch of high quality, all reflect superb
New Function Rich configuration
Program and Structure optimization
LC-100 HPLC Gradient System Flow rate range: 0.001-10.00 mL/min, in 0.001 mL increments
Flow accuracy: ≤±0.1% (1.00ml/min, H2O,R.T.,8.5MPa);
Flow precision: RSD≤0.06% (1.00ml/min, H2O,R.T.,8.5MPa)
Maximum pressure: 6090 psi ; Pressure accuracy: ≤±1.0%;
Pressure pulsation: ≤0.1 MPa P100 C Dual-piston reciprocating parallel Binary High Pressure Gradient System
Flow Presion: 0RSD<0.06%; Fluctuation: 0.1MPa ;Mixed Liquid: 2 to 4 kinds of liquid ; Mixing Principle :Gradient proportion of the switching time control of proportional valve ; Mixing Accuracy :±1%;Mixing Precision:±0.2%; Power Supply: 110V / 220V EX1600 SM High pressure gradient mixer Mixed volume :750µL ; Mixed Liquid: 2 kinds of liquid ; UV-100 Detector : Wavelength Range: 190-700 nm
Flow cell volume: 8nm; Wavelength Accuracy: ±1 nm ;
Wavelength Precision: <0.1nm;
Minimum Detectable Concentration: 3×10-9 g/mL; Noise :±0.25×10-5AU(Static)/±1×10-5AU(dynamic, under specified requirements);
Drift :≤0.4×10-4AU/h(Static)/±2×10-4AU/h (dynamic, under specified requirements); Power Supply: 110V / 220V ; WS-100 Chromatographic Data-processing Workstation
Signal Transmission Form :Full digital signal;
Operation Mode:Reflecting and monitoring the operation of instruments’ each unit (editing gradient program for changing wavelength scanning, setting event procedure, spectrum scanning, spectrogram comparison, etc.)
Original Series C18 5μm ID4.6×250mm